Measurements and Math Functions 6
InfiniiVision Oscilloscopes User’s Guide 215
To make cursor measurements
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Press the [Cursors] key. The key will illuminate, indicating that cursors are “on”.
(Press the key again when you want to turn cursors off.)
3 In the Cursors Menu, press Mode; the, select the desired mode:
ManualX, 1/X, and Y values are displayed. X is the difference between
the X1 and X2 cursors and Y is the difference between the Y1 and Y2 cursors.
Track Waveform — As you move a marker horizontally, the vertical amplitude of
the waveform is tracked and measured. At the bottom of the display, the time and
voltage positions are shown for the markers. The vertical (Y) and horizontal (X)
differences between the markers are shown as X and Y values.
Binary — Logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed above the softkeys in binary. The display is color coded to
match the color of the related channel’s waveform.
Hex — Logic levels of displayed waveforms at the current X1 and X2 cursor
positions are displayed above the softkeys in hexadecimal.
NOTE
If you intend to make cursor measurements on a trace that you recall from memory, be
sure to recall both setup and trace. See “To recall waveform trace and/or oscilloscope
setup” on page 256.
Cursor X1 values
for analog channels
1 through 4
Cursor X1 values for
digital channels
D15 Through D0