Agilent Technologies 54502A Manuel de service Page 40

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 111
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 39
2. Adjust pulse generator for the output in figure 3-3.
Setting for HP 8161A:
Parameter Output A Output B
Input Mode Norm ---
Period (PER) 60 ns ---
Width (WID) 10 ns 10 ns
Leading Edge (LEE) 1 ns 1 ns
Trailing Edge (TRE) 1 ns 1 ns
High Level (HIL) 3.2 V 3.2 V
Low Level (LOL) 0 V 0 V
Delay (DEL) 0 ns 0 ns
Output Mode ENABLE ENABLE
3. Assign the pod under test to Analyzer 1 in the System Configuration as in
3-4. Refer to steps a through c if unfamiliar with menu.
a. Move cursor to Type field of Analyzer 1 and SELECT.
b. Set the analyzer Type to State using cursor and SELECT.
c. Move cursor to Pod to be tested and assign to Machine 1.
Figure 3-3. Waveform for Data Test 1
Figure 3-4. System Configuration for Data Test 1
HP 1650B/1651B Performance Tests
Service Manual 3-5