Agilent Technologies 54502A Manuel de service Page 55

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 111
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 54
a. Move cursor to Clock field and SELECT, then assign Mixed Clocks.
b. Assign falling transition of the J clock to Master Clock and rising
transition of the J clock to Slave Clock.
c. Assign channels 0-3 and 8-11 of the pod under test.
d. Set Clock Period to > 60 ns.
5. Set the State Trace Specification without sequencing levels and Count Off
as in previoius figure 3-13.
6. Press RUN. The State Listing displays alternating Fs and 0s for the channels
under test as in figure 3-24.
Note To ensure consistent pattern of alternating Fs and 0s, use the front-panel ROLL
field and knob to scroll through State Listing.
7. Connect the next clock to the test connector and repeat steps 4 and 6.
Repeat until all clocks have been tested (clocks J, K, L, M and N).
8. Remove the probe tip assembly from the logic analyzer probe cable and
attach to the next logic analyzer probe cable to be tested. Take care not to
dislodge grabbers from the test connector.
9. Repeat steps 3, 4, 6 and 7 until all pods have been tested (pods 1 through 5).
Start again with the falling edge of J clock as the Master and rising edge of
the J clock as the Slave clock. Repeat until channels 0 - 3 and 8 - 11 of all
pods have been tested (pods 1 - 5).
10. Disconnect bits 0-3 and bits 8-11 from the test connector. Attach bits 4-7 and
bits 12-15 to the test connector. Repeat steps 3, 4, 6, 7 and 8 until all pods
have been tested (pods 1 through 5).
Figure 3-24. State Listing for Data Test 5
Performance Tests HP 1650B/1651B
3-20 Service Manual
Vue de la page 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 110 111

Commentaires sur ces manuels

Pas de commentaire