5. Adjust the power supply output for + 150 mV.
6. Press RUN. Data displayed on Timing Waveforms is be high for the pod and
channels under test as in figure 3-38.
7. Adjust power supply output for -150 mV.
8. Press RUN. Data displayed on Timing Waveforms is all low for the
channels under test as in figure 3-39.
9. Return to Timing Format Specification and change User Defined Pod
Threshold to + 9.9 V.
10. Adjust power supply for output of + 10.2 V.
11. Press RUN. Data displayed on Timing Waveforms all high for the pod and
channels under test as in previous test figure 3-37.
12. Adjust power supply for output of + 9.6 V.
13. Press RUN. Data displayed on Timing Waveforms is all low for the pod and
channels under test as in previous test figure 3-40.
Figure 3-38. Threshold Accuracy Timing Waveforms 1
Figure 3-39. Threshold Accuracy Timing Waveforms 2
Performance Tests HP 1650B/1651B
3-30 Service Manual