Threshold
Accuracy Test
Description: This performance test verifies the threshold accuracy within the
three ranges stated in the specification.
Specification:
Threshold accuracy: 150 mV accuracy over the range -2.0 to + 2.0 volts; 300 mV
accuracy over the ranges -9.9 to -2.1 volts and + 2.1 to + 9.9 volts.
Equipment:
Power Supply .................................... HP 6216B
Test Connector (1)......................... see figure 3-1
Procedure:
1. Connect the test equipment as in figure 3-35.
Note In this setup, only eight channels are tested at a time to minimize loading. Ground
lead must be grounded to ensure accurate test results.
2. Assign the pod under test to Analyzer 1 in System Configuration as in
previous figure 3-31.
3. Configure Timing Format Specification for User Defined pod threshold of
0.0 V for the pod under test and assign the lower eight bits in the bit
assignment field as in figure 3-36. Refer to steps a and b if unfamiliar with
menu.
a. Move cursor to Pod Threshold field and SELECT. Move cursor to
User-defined field and SELECT. Enter the appropriate voltage
threshold.
Figure 3-35. Threshold Accuracy Test Setup
Performance Tests HP 1650B/1651B
3-28 Service Manual