Agilent B2900 SCPI Command Reference, Edition 2 3- 9
Common Commands
*OPC
*OPC
This command starts to monitor pending operations, and sets/clears the operation
complete (OPC) bit in the Standard Event Status register as follows.
If there is no pending operation, the OPC bit is set to 1.
If there are any pending operations, the OPC bit is set to 0. The bit will be set to
1 again when all pending operations are completed.
The *OPC command is required to enable the OPC bit. To stop monitoring pending
operations (disable OPC bit), execute the
*CLS command.
Other commands cannot be executed until this command completes.
Syntax *OPC
*OPC?
Query response 1 <newline><^END>
The query returns 1 if the instrument has completed all pending operations sent
before this command. Response data type is NR1.
See Also *WAI