Agilent B2900 SCPI Command Reference, Edition 2 4- 79
Subsystem Commands
:SENSe:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:MODE
:SENSe:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AU
TO:MODE
Selects the operation mode of the automatic measurement ranging. This command
setting is not effective if the automatic ranging is disabled by the
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:RANGe:AUTO command.
Syntax :SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:MODE mode
:SENSe[c]:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:MODE?
For <CURRent[:DC]|VOLTage[:DC]>, specify CURRent[:DC] for current
measurement, or VOLTage[:DC] for voltage measurement.
Parameter mode NORMal (default)|RESolution|SPEed. Parameter data type is
CPD.
mode=NORMal supports basic operation and downward changing operation
described below
mode=RESolution supports basic operation and upward changing operation
described below
mode=SPEed supports basic operation and upward and downward changing
operations described below
Basic operation
The channel automatically sets the range which provides the best resolution to
perform the measurement.
Upward changing operation
If measured data value1, the range shifts upward after a measurement.
value1 = measurement range rate / 100
Downward changing operation
If measured data value2, the range shifts downward immediately.
value2 = measurement range rate / 1000
rate value is set by the
:SENSe:<CURRent[:DC]|VOLTage[:DC]>:RANGe:AUTO:THReshold command.
Query response mode <newline>