Agilent B2900 SCPI Command Reference, Edition 2 4- 91
Subsystem Commands
:SENSe:WAIT[:STATe]
:SENSe:WAIT[:STATe]
Enables or disables the measurement wait time for the specified channel. The wait
time is defined as the time the measurement channel cannot start measurement after
the start of a DC output or the trailing edge of a pulse.
Syntax :SENSe[c]:WAIT[:STATe] mode
:SENSe[c]:WAIT[:STATe]?
Parameter mode 0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the measurement wait time. The wait time is set to 0.
mode=1 or ON enables the measurement wait time given by the following formula.
:SENSe:WAIT:AUTO ON|1 condition:
wait time = gain initial wait time + offset
:SENSe:WAIT:AUTO OFF|0 condition:
wait time = offset
The initial wait time is automatically set by the instrument and cannot be changed.
gain and offset are set by the :SENSe:WAIT:GAIN and :SENSe:WAIT:OFFSet
commands, respectively.
Query response mode <newline>
mode is 0 or 1, and indicates that the measurement wait time is disabled or enabled,
respectively. Response data type is NR1.
Example :SENS:WAIT 0
:SENS2:WAIT:STAT?