Agilent B2900 SCPI Command Reference, Edition 2 5- 17
Error Messages
B2900 Specific Error
Error 356 Emergency; Sense data FIFO overflow detected; Channel n
Sense data FIFO overflow was detected in channel n. All channel output is changed
to 0 V and the all output switch is opened. Execute the *TST? command.
Error 361 Emergency; Internal communication failure detected by
CPU; Channel n
Channel n internal communication failure was detected by CPU. All channel output
is changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 362 Emergency; Internal command queue overflow detected;
Channel n
Internal command queue overflow was detected in channel n. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 363 Emergency; Sense data was not received for acquire
trigger; Channel n
Channel n sense data was not received for acquire trigger. All channel output is
changed to 0 V and the all output switch is opened. Execute the *TST? command.
Error 364 Emergency; Unexpected sense data was received; Channel n
Unexpected sense data was received. All channel output is changed to 0 V and the
all output switch is opened. Execute the *TST? command.
Error 365 Emergency; Sense data was not received in Timer period;
Channel n
Data communication failure. All channel output is changed to 0 V and the all output
switch is opened. Execute the *TST? command.
Error 366 Emergency; Timestamp FIFO overflow detected; Channel n
Data communication failure. All channel output is changed to 0 V and the all output
switch is opened. Execute the *TST? command.
Error 700 ProgramMemory; Program size overflow
Program memory cannot save the program. Reduce program size.
Error 701 ProgramMemory; Invalid variable
Appropriate variable name must be specified.