Agilent Technologies E5503A Guide de l'utilisateur Page 123

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 465
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 122
Agilent Technologies E5500 Phase Noise Measurement System
6-21
Absolute Measurement Fundamentals
Evaluating Noise Above the Small Angle Line
Figure 6-11 Graph of Phase Lock Loop Bandwidth Provided by the Peak Tuning
Range
1. Once the beatnote is displayed, press the press [[RANGE ]] , press
[[AUTO RANGE OFF ]] , and press [[SINGLE AUTO RANGE ]] on
the RF analyzer.
2. Set the span width on the RF analyzer to approximately 4 x PLL
bandwidth. Adjust the beatnote to position it near the center of the
display.
NOTE If you are not able to tune the beatnote to 2 X PLL bandwidth (center of
display) due to frequency drift, refer to Tracking Frequency Drift in this
section for information about measuring drifting signals. If you are able to
locate the beatnote, but it distorts and then disappears as you adjust it
towards 0 Hz, then your sources are injection locking to each other. Set the
beatnote to the lowest frequency possible before injection locking occurs
and then refer to !!Minimizing Injection Locking!! in this section for
recommended actions.
3. Press the [[ AVG ]] key, and then the RMS key. Wait for the trace to
return and then press [[ MKR ]] and MKR to Peak.
4. Press [[ REL MKR ]] , and MKR REF.
5. Press the [[ DEFINE TRACE ]]press the [[ and the MATH FUNCTION
keys.
Vue de la page 122
1 2 ... 118 119 120 121 122 123 124 125 126 127 128 ... 464 465

Commentaires sur ces manuels

Pas de commentaire