Agilent Technologies E5503A Guide de l'utilisateur Page 304

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 465
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 303
Agilent Technologies E5500 Phase Noise Measurement System
11-15
FM Discriminator Measurement Examples
FM Discriminator Measurement using Double-Sided Spur Calibration
The segment data will be displayed on the computer screen as the data is
taken until all segments have been taken over the entire range you specified
in the Measurement definition’s Type and Range.
When the
Measurement is
Complete
When the measurement is complete, refer to Chapter 15, “Evaluating Your
Measurement Results” for help in evaluating your measurement results. (If
the test system has problems completing the measurement, it will inform you
by placing a message on the computer display.
Figure 11-3 on page 11-15 shows a typical absolute measurement using FM
discrimination.
Figure 11-3 Typical Phase Noise Curve using Double-Sided Spur Calibration.
Vue de la page 303
1 2 ... 299 300 301 302 303 304 305 306 307 308 309 ... 464 465

Commentaires sur ces manuels

Pas de commentaire