Agilent Technologies E5503A Guide de l'utilisateur Page 281

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 465
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 280
Agilent Technologies E5500 Phase Noise Measurement System
9-13
Residual Measurement Examples
Amplifier Measurement Example
4. The system software will then prompt you to set the phase noise
software’s meter to quadrature by adjusting the phase shifter until the
meter indicates 0 volts, then press Continue.
5. The system will now measure the noise data.
The system can now run the measurement. The segment data will be
displayed on the computer screen as the data is taken until all segments have
been taken over the entire range you specified in the Measurement
definition’s Type and Range.
When the
Measurement is
Complete
When the measurement is complete, refer to Chapter 15, “Evaluating Your
Measurement Results” for help in evaluating your measurement results.
Figure 9-3 on page 9-14 shows a typical phase noise curve for an RF
Amplifier.
Vue de la page 280
1 2 ... 276 277 278 279 280 281 282 283 284 285 286 ... 464 465

Commentaires sur ces manuels

Pas de commentaire