Agilent Technologies E5503A Guide de l'utilisateur Page 356

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 465
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 355
14-6
Agilent Technologies E5500 Phase Noise Measurement System
Baseband Noise Measurement Examples
Baseband Noise without using a Test Set Measurement Example
Baseband Noise without using a Test Set Measurement
Example
This measurement example will help you measure the noise voltage of a
source.
NOTE To ensure accurate measurements, you should allow the UUT and
measurement equipment to warm up at least one hour before making the
noise measurement.
Defining the
Measurement
1. From the File menu, choose Open.
2. If necessary, choose the drive or directory where the file you want is
stored.
3. In the File Name box, choose “BBnoise_without_testset_89410.pnm”.
Vue de la page 355
1 2 ... 351 352 353 354 355 356 357 358 359 360 361 ... 464 465

Commentaires sur ces manuels

Pas de commentaire