Agilent Technologies E5500 Phase Noise Measurement System
18-7
Reference Graphs and Tables
AM Calibration
AM Calibration
The AM detector sensitivity graph shown above is used to determine the
equivalent phase Detector Constant from the measured AM Detector input
level or from the diode detector's dc voltage. The equivalent phase Detector
Constant (phase slope) is read from the left side of the graph while the
approximate detector input power is read from the right side of the graph.