Agilent Technologies E5500 Phase Noise Measurement System
12-9
AM Noise Measurement Fundamentals
Method 1: User Entry of Phase Detector Constant
Figure 12-7 AM Noise Calibration Setup
3. Locate the drive level on the AM sensitivity graph (figure 3-5), and enter
the data.
4. Measure the noise data and interpret the results. The measured data will
be plotted as single-sideband AM noise in dBc/Hz.
NOTE The quadrature meter should be at zero volts due to the blocking capacitor at
the AM detectors output.
Figure 12-8 AM Detector Sensitivity Graph