Agilent Technologies E5500 Phase Noise Measurement System
7-77
Absolute Measurement Examples
RF Synthesizer using EFC
Selecting a Reference
Source
1. From the Define menu, choose Measurement; then choose the Block
Diagram tab from the Define Measurement window.
2. From the Reference Source pull-down list, select your reference source.
3. When you have completed these operations, click the Close button
.
Setup Considerations
for the RF Synthesizer
using EFC
Measurement
Measurement Noise Floor
The signal amplitude at the R input (Signal Input) port on the
Agilent/HP 70420A sets the measurement noise floor level. Use the
following graph to determine the amplitude required to provide a noise floor
level that is below the expected noise floor of your UUT.