Agilent Technologies 85032B/E 50 Spécifications Page 115

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 269
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 114
Adjustments 8712ET/ES and 8714ET/ES
B* Amplitude Correction, Adjustment #110 RF Network Analyzers
2-22 Service Guide
Procedure
1. Press
.
2. This is a fully automated adjustment handled by the internal firmware. Follow the
instructions on the analyzer's display.
The analyzer presets after the test is done.
3. If you are not going to make any more adjustments, press
.
a. Insert a formatted disk into the internal disk drive.
b. Press .
This creates a file (or writes over an existing file) where the correction constants are
stored.
PRESET SYSTEM OPTIONS
Service
Test and Adjustments
Select Adjustment
110 ENTER
Execute Test
Service
Update Corr Const
Store CC to EPROM
SYSTEM OPTIONS
Service
Update Corr Const
Store CC to Disk
Vue de la page 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 268 269

Commentaires sur ces manuels

Pas de commentaire