Agilent Technologies 85032B/E 50 Spécifications Page 131

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 269
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 130
8712ET/ES and 8714ET/ES Assembly Replacement
RF Network Analyzers A5 Receiver Assembly Component Replacement
Service Guide 3-13
A5 Receiver Assembly Component Replacement
The A5 receiver board isnotintended to be replaced at the assembly level. The receiver
assembly can be troubleshot down to module or component level. Instructions for replacing
the module and components are included in this section.
CAUTION Perform the followingonly at an approved ESD workstation. SeeFigure 4-1
on page4-3.
NOTE When connecting an SMA adapter (see item 2 inTable 7-8,Table 7-9, and
Table 7-10), use the 7/32-inch wrench to hold the adapter in place while
tightening the attaching cables.
Removing the Receiver Assembly
1.Disconnect the power cord.
2.Disconnect any cables from the EXT DET X-input, EXT DET Y-input, and AUX INPUT
rear panel BNC connectors.
3.Remove nuts and washers from the AUX INPUT BNC connector.
4.Remove the three anchoring screws on the bottom of the analyzer. SeeFigure 3-6.
Figure 3-6. Location of the A5 Receiver Anchoring Screws
Vue de la page 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 268 269

Commentaires sur ces manuels

Pas de commentaire