Agilent Technologies 85032B/E 50 Spécifications Page 145

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 269
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 144
8712ET/ES and 8714ET/ES Troubleshooting and Block Diagrams
RF Network Analyzers Introduction
Service Guide 4-3
Figure 4-1 Example of an Antistatic Workstation
For disassembly procedures, see Chapter 3.
WARNING Some parts in the analyzer have sharp edges. Work carefully to avoid
injury.
Before replacing an assembly, inspect the board for obvious, easy-to-fix defects.
Examples include:
bent pins on ICs
no solder in holes around the edges of shields
cold solder joints
Service Kit
The service kit (08712-60012) for the 871x series analyzers contains several parts that are
useful for adjusting and servicing the analyzer. The following parts are included:
extender board assembly that allows you to operate selected internal board assemblies
outside the analyzer chassis
GPIB connector extender that allows you to mate an GPIB cable when the extender
board assembly is attached to the analyzer
two SMB extension cables, 12 inches long (30-cm)
modified 5/16-inch wrench for better access to SMA cable nuts
IC extractor tool for removing ICs such as the U335 bootROM
6-inch (150-mm) semirigid coaxial extension cable
Vue de la page 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 268 269

Commentaires sur ces manuels