Service Related Menus 8712ET/ES and 8714ET/ES
GPIB Command Reference for Service RF Network Analyzers
5-26 Service Guide
TEST:STATe <string> CHAR Selects the state of the active
adjustment or self-test. Choose
from RUN|CONTinue|STOP.
Query response will be
NULL|PAUS|DONE.
TEST:VALue <num> NR1 Sets or queries a value for an
adjustment or self-test.
*TST? Query
NR1
Returns the result of a complete
selftest. An ASCII 0 (zero)
indicates no failures found. Any
other character indicates a failure.
Command Form Description