8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers System Performance
Service Guide 1-49
Table 1-5 System Performance, Corrected, 75, T/R Calibration
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25°±5 °C, with less than 1 °C deviation from
the calibration temperature.
8712ES/8714ES (75 )
85036B/E (Type-N, 75 ) Cal Kit, T/R Calibration
Description
Specification
1
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 48 43
Source Match:
Reflection (One-Port Cal) 41 35
Transmission (Enhanced Response Cal) 41 35
Transmission (Response Cal) 18 15
Load Match 18 15
Reflection Tracking ±0.021 ±0.02
Transmission Tracking:
Enhanced Response Cal ±0.042 ±0.062
Response Cal ±0.17 ±0.3
Vue de la page 56