8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Test Port Input
Service Guide 1-65
Table 1-9 Test Port Input (continued)
1. The System Dynamic Range is calculated as the difference between
the receiver noise floor and the minimum of either the source maxi-
mum output (maximum power setting minus output power level
accuracy) or the receiver maximum input. System Dynamic Range
applies to transmission measurements only, since reflection mea-
surements are limited by directivity. The System Dynamic Range for
8714ET/ES 75 analyzers is not a specification for frequencies
>2 GHz; it is a characteristic. For the 8712ES/8714ES, external
broadband detectors will provide more dynamic range than the
internal broadband detectors.
2. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
8712ET/ES and 8714ET/ES Test Port Input
Description
Spec.
(in dB)
Typical
(in dB)
System Bandwidths:
Fine
(15 Hz)
Fine
(15 Hz)
Med Wide
(4000 Hz)
Wide
(6500 Hz)
System Dynamic Range
1
8712ET:
50 :
With Attenuator Option 1E1
2
:
Narrowband 115 121 96 58
Broadband (Internal) 60 67 60 44
Without Attenuator Option 1E1
Narrowband 115 121 96 58
Broadband (Internal) 62 68 61 45
75 :
With Attenuator Option 1E1
2
:
Narrowband 110 118 93 55
Broadband (Internal) 53 61 54 39
Without Attenuator Option 1E1
Narrowband 113 119 94 56
Broadband (Internal) 56 62 55 40