Module 5
Basic Measurement
5-23
Two SMUs in Series
VAR1
VAR1'
SMU1
V=0 to 100
SMU2
V=0 to -100
200 V Resistor Sweep
Base
Collector
Const.
Emitter
VAR1'
VAR1
SMU1
I=0
SMU2
V=0 to -100
SMU3
V=0 to 100
200 V BVceo
Base current = 0
or leave open
V1
V2
WARNING
200 Volt Sweep (400 Volt with 2 HPSMUs)
Higher voltages are necessary to observe breakdown conditions, such as BVceo. In the case of
BVceo the base must be open (no SMU) or the SMU must be set as a current source with I=0.
One SMU must sweep in a positive direction while the other sweeps in a negative direction. The
result is a floating measurement with double the differential voltage. A resistor is used in the next
few slides to demonstrate a 200 V sweep.
Note: Using two HPSMUs, it is possible to sweep a total of 400 Volts.
Commentaires sur ces manuels