Module 6
Low Current Measurement
6-14
Linear, with zero set to mid scale
ZERO CHECK
Display Setup
The best check is done with a LINEAR scale which brackets either side of ZERO by 100 fA or less.
A log scale is particularly unacceptable, Log plots of zero are not possible, and all readings of a log
plot must be either all positive or all negative.