:BUS<n> Commands 8
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 199
:BUS<n>:BIT<m>
(see page 790)
Command Syntax
:BUS<n>:BIT<m> <display>
<display> ::= {{1 | ON} | {0 | OFF}}
<n> ::= An integer, 1 or 2, is attached as a suffix to BUS
and defines the bus that is affected by the command.
<m> ::= An integer, 0,..,7, is attached as a suffix to BIT
and defines the digital channel that is affected by the command.
The :BUS<n>:BIT<m> command includes or excludes the selected bit as
part of the definition for the selected bus. If the parameter is a 1 (ON),
the bit is included in the definition. If the parameter is a 0 (OFF), the bit
is excluded from the definition. Note: BIT0- 7 correspond to DIGital0- 7.
Query Syntax
:BUS<n>:BIT<m>?
The :BUS<n>:BIT<m>? query returns the value indicating whether the
specified bit is included or excluded from the specified bus definition.
Return Format
<display><NL>
<display> ::= {0 | 1}
See Also "Introduction to :BUS<n> Commands" on page 198
":BUS<n>:BITS" on page 200
":BUS<n>:CLEar" on page 202
":BUS<n>:DISPlay" on page 203
":BUS<n>:LABel" on page 204
":BUS<n>:MASK" on page 205
Example Code
' Include digital channel 1 in bus 1:
myScope.WriteString ":BUS1:BIT1 ON"