Agilent Technologies 2000 X-Series Manuel de service Page 595

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 930
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 594
:TRIGger Commands 29
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 595
:TRIGger:LEVel:ASETup
(see page 790)
Command Syntax
:TRIGger:LEVel:ASETup
The :TRIGger:LEVel:ASETup command automatically sets the trigger levels
of all displayed analog channels to their waveforms' 50% values.
If AC coupling is used, the trigger levels are set to 0 V.
When High and Low (dual) trigger levels are used (as with Rise/Fall Time
and Runt triggers, for example), this command has no effect.
See Also ":TRIGger[:EDGE]:LEVel" on page 603