556 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
26 :SEARch Commands
:SEARch:SERial:SPI:MODE
(see page 790)
Command Syntax
:SEARch:SERial:SPI:MODE <value>
<value> ::= {MOSI | MISO}
The :SEARch:SERial:SPI:MODE command specifies whether the SPI search
will be on the MOSI data or the MISO data.
Data values are specified using the :SEARch:SERial:SPI:PATTern:DATA
command.
Query Syntax
:SEARch:SERial:SPI:MODE?
The :SEARch:SERial:SPI:MODE? query returns the current SPI search mode
setting.
Return Format
<value><NL>
<value> ::= {MOSI | MISO}
See Also Chapter 26, “:SEARch Commands,” starting on page 531
":SEARch:SERial:SPI:PATTern:DATA" on page 557
Vue de la page 555