Agilent Technologies 2000 X-Series Manuel de service Page 268

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 930
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 267
268 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
14 :DVM Commands
:DVM:MODE
(see page 790)
Command Syntax
:DVM:MODE <dvm_mode>
<dvm_mode> ::= {ACRMs | DC | DCRMs | FREQuency}
The :DVM:MODE command sets the digital voltmenter (DVM) mode:
ACRMs — displays the root-mean-square value of the acquired data,
with the DC component removed.
DC — displays the DC value of the acquired data.
DCRMs — displays the root-mean-square value of the acquired data.
FREQuency — displays the frequency counter measurement.
Query Syntax
:DVM:MODE?
The :DVM:MODE? query returns the selected DVM mode.
Return Format
<dvm_mode><NL>
<dvm_mode> ::= {ACRM | DC | DCRM | FREQ}
See Also ":DVM:ENABle" on page 266
":DVM:SOURce" on page 269
":DVM:ARANge" on page 264
":DVM:CURRent" on page 265
":DVM:FREQuency" on page 267
Vue de la page 267
1 2 ... 263 264 265 266 267 268 269 270 271 272 273 ... 929 930

Commentaires sur ces manuels

Pas de commentaire