Agilent Technologies 2000 X-Series Manuel de service Page 468

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 930
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 467
468 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE
(see page 790)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE <value>
<value> ::= {STANdard | EXTended}
The :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE command sets the CAN
identifier mode. STANdard selects the standard 11-bit identifier. EXTended
selects the extended 29- bit identifier. The CAN identifier is set by the
:SBUS<n>:CAN:TRIGger:PATTern:ID command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE?
The :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE? query returns the current
setting of the CAN identifier mode.
Return Format
<value><NL>
<value> ::= {STAN | EXT}
Errors "- 241, Hardware missing" on page 753
See Also "Introduction to :TRIGger Commands" on page 589
":SBUS<n>:MODE" on page 451
":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 465
":SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth" on page 466
":SBUS<n>:CAN:TRIGger:PATTern:ID" on page 467
Vue de la page 467
1 2 ... 463 464 465 466 467 468 469 470 471 472 473 ... 929 930