678 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
31 :WGEN Commands
:WGEN:MODulation:FUNCtion
(see page 790)
Command Syntax
:WGEN:MODulation:FUNCtion <shape>
<shape> ::= {SINusoid | SQUare| RAMP}
The :WGEN:MODulation:FUNCtion command specifies the shape of the
modulating signal.
When the RAMP shape is selected, you can specify the amount of time per
cycle that the ramp waveform is rising with the
:WGEN:MODulation:FUNCtion:RAMP:SYMMetry command.
This command applies to AM and FM modulation. (The FSK modulation
signal is a square wave shape.)
Query Syntax
:WGEN:MODulation:FUNCtion?
The :WGEN:MODulation:FUNCtion? query returns the specified modulating
signal shape.
Return Format
<shape><NL>
<shape> ::= {SIN | SQU| RAMP}
See Also ":WGEN:MODulation:AM:DEPTh" on page 672
":WGEN:MODulation:AM:FREQuency" on page 673
":WGEN:MODulation:FM:DEViation" on page 674
":WGEN:MODulation:FM:FREQuency" on page 675
":WGEN:MODulation:FSKey:FREQuency" on page 676
":WGEN:MODulation:FSKey:RATE" on page 677
":WGEN:MODulation:FUNCtion:RAMP:SYMMetry" on page 679
":WGEN:MODulation:STATe" on page 681
":WGEN:MODulation:TYPE" on page 682