Agilent Technologies 2000 X-Series Manuel de service Page 213

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 930
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 212
:CALibrate Commands 9
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 213
:CALibrate:STARt
(see page 790)
Command Syntax
:CALibrate:STARt
The CALibrate:STARt command starts the user calibration procedure.
See Also "Introduction to :CALibrate Commands" on page 207
":CALibrate:PROTected" on page 212
NOTE
Before starting the user calibration procedure, you must set the rear panel CALIBRATION
switch to UNPROTECTED, and you must connect BNC cables from the TRIG OUT connector
to the analog channel inputs. See the User's Guide for details.
Vue de la page 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 929 930

Commentaires sur ces manuels

Pas de commentaire