Agilent Technologies 2000 X-Series Manuel de service Page 730

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 930
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 729
730 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
33 Obsolete and Discontinued Commands
:MEASure:TSTOp
(see page 790)
Command Syntax
:MEASure:TSTOp <value> [suffix]
<value> ::= time at the stop marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTOp?
The :MEASure:TSTOp? query returns the time at the stop marker (X2
cursor).
Return Format
<value><NL>
<value> ::= time at the stop marker in NR3 format
See Also "Introduction to :MARKer Commands" on page 318
"Introduction to :MEASure Commands" on page 340
":MARKer:X1Position" on page 320
":MARKer:X2Position" on page 322
":MARKer:XDELta" on page 324
":MEASure:TDELta" on page 727
":MEASure:TSTArt" on page 729
NOTE
The short form of this command, TSTO, does not follow the defined Long Form to Short
Form Truncation Rules (see page 792). The normal short form "TST" would be the same for
both TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.
NOTE
The :MEASure:TSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X2Position command (see page 322) instead.
Vue de la page 729
1 2 ... 725 726 727 728 729 730 731 732 733 734 735 ... 929 930

Commentaires sur ces manuels

Pas de commentaire