Agilent Technologies InfiniiVision 6000 Series Manuel de service Page 145

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 934
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 144
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 145
Query Syntax *SRE?
The *SRE? query returns the current value of the Service Request Enable
Register.
Return Format
<mask><NL>
<mask> ::= sum of all bits that are set, 0,..,255;
an integer in NR1 format
See Also "Introduction to Common (*) Commands" on page 128
"*STB (Read Status Byte)" on page 146
"*CLS (Clear Status)" on page 129
5 ESB Event Status Bit Interrupts when enabled conditions in the
Standard Event Status Register (ESR) occur.
4 MAV Message Available Interrupts when messages are in the Output
Queue.
3 --- --- (Not used.)
2 MSG Message Interrupts when an advisory has been displayed
on the oscilloscope.
1 USR User Event Interrupts when enabled user event conditions
occur.
0 TRG Trigger Interrupts when a trigger occurs.
Table 43 Service Request Enable Register (SRE) (continued)
Bit Name Description When Set (1 = High = True), Enables:
Vue de la page 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 933 934

Commentaires sur ces manuels

Pas de commentaire