Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 661
:WAVeform:YINCrement
(see page 798)
Query Syntax
:WAVeform:YINCrement?
The :WAVeform:YINCrement? query returns the y- increment value in volts
for the currently specified source. This value is the voltage difference
between consecutive data values. The y- increment for digital waveforms is
always "1".
Return Format
<value><NL>
<value> ::= y-increment value in the current preamble in 32-bit
floating point NR3 format
See Also "Introduction to :WAVeform Commands" on page 630
":WAVeform:PREamble" on page 645
Example Code "Example Code" on page 646