518 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:FLEXray:AUTosetup
(see page 798)
Command Syntax
:TRIGger:FLEXray:AUTosetup
The :TRIGger:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
Sets the selected source channel's impedance to 50 Ohms.
Sets the selected source channel’s probe attenuation to 10:1.
Sets the trigger level (on the selected source channel) to - 300 mV.
Turns on trigger Noise Reject.
Turns on Serial Decode.
Sets the trigger type to FlexRay.
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:FLEXray:TRIGger" on page 528
":TRIGger:FLEXray:BAUDrate" on page 519
":TRIGger[:EDGE]:LEVel" on page 513
":TRIGger:FLEXray:SOURce" on page 527
NOTE
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.