620 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:UART:SOURce:TX
(see page 798)
Command Syntax
:TRIGger:UART:SOURce:TX <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for the MSO models
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:UART:SOURce:TX command controls which signal is used as
the Tx source by the serial decoder and/or trigger when in UART mode.
Query Syntax
:TRIGger:UART:SOURce:TX?
The :TRIGger:UART:SOURce:TX? query returns the current source for the
UART Tx signal.
Return Format
<source><NL>
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:UART:TYPE" on page 621
":TRIGger:UART:BITorder" on page 612