Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 359
:MEASure:VAVerage
(see page 798)
Command Syntax
:MEASure:VAVerage [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VAVerage command installs a screen measurement and
starts an average value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VAVerage? [<source>]
The :MEASure:VAVerage? query returns the average value of an integral
number of periods of the signal. If at least three edges are not present, the
oscilloscope averages all data points.
Return Format
<value><NL>
<value> ::= calculated average value in NR3 format
See Also "Introduction to :MEASure Commands" on page 324
":MEASure:SOURce" on page 349