402 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MTESt:SOURce
(see page 798)
Command Syntax
:MTESt:SOURce <source>
<source> ::= CHANnel<n>
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MTESt:SOURce command selects the channel which is configured by
the commands contained in a mask file when it is loaded.
Query Syntax
:MTESt:SOURce?
The :MTESt:SOURce? query returns the channel which is configured by the
commands contained in the current mask file.
Return Format
<source><NL>
<source> ::= {CHAN<n> | NONE}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
See Also "Introduction to :MTESt Commands" on page 372