622 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:UART:WIDTh
(see page 798)
Command Syntax
:TRIGger:UART:WIDTh <width>
<width> ::= {5|6|7|8|9}
The :TRIGger:UART:WIDTh command determines the number of bits (5- 9)
for each message "byte" for the serial decoder and/or trigger when in
UART mode.
Query Syntax
:TRIGger:UART:WIDTh?
The :TRIGger:UART:WIDTh? query returns the current UART width setting.
Return Format
<width><NL>
<width> ::= {5|6|7|8|9}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:UART:TYPE" on page 621