424 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:SAVE:LISTer[:STARt]
(see page 798)
Command Syntax
:SAVE:LISTer[:STARt] [<file_name>]
<file_name> ::= quoted ASCII string
The :SAVE:LISTer[:STARt] command saves the Lister display data to a file.
See Also "Introduction to :SAVE Commands" on page 416
":SAVE:FILename" on page 417
":LISTer Commands" on page 304
NOTE
If a file extension is provided as part of a specified <file_name>, it must be ".csv".