DIGital Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 33
NOTE This parameter has no influence if the delay control input for the
channel is switched on (
:PM[1|2] ON
).
The figure below shows the normal and complement output with
the crossover point set to 50% and 70% respectively.
Parameter Numeric values (in %) in the range of 20 … 80.
*RST value 50
Example Set the variable crossover point to 70%.
:DIG:SIGN:CROS 70
50 %
Normal
Out
50 %
Compl.
Out
70 %
Compl.
Out
Normal
Out
70 %