SOURce Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 63
[:SOURce]:VOLTage[0|1|2][:LEVel]
[:IMMediate]:LOW
Syntax
[:SOUR]:VOLT[0|1|2][:LEV][:IMM]:LOW[?]
Form Set & Query
Description Programs the low- level of the output signal for the trigger output
and the channels.
Parameter Numeric [uV|mV|V] |MIN|MAX
Value coupling Amplitude = High – Low
Offset = (High – Low) / 2
Range coupling High- level
*RST value Trigger channel (0): 0.0 V
Channels 1 and 2: 0.0 V
Specified limits Trigger channel (0): –2.0 … +2.95 V
Channels 1 and 2: –2.0 … +2.95 V
Absolute limits Trigger output (0):2.2 +3.2 V
Channels 1 and 2: –2.2 … +3.2 V
Example Set Channel 1 low- level to –1 V.
:VOLT1:LOW -1V