SCPI Commands Reference Status Handling Commands
70 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
:STATus:OPERation
This command tree accesses the OPERation status group.
The OPERation status group is not used by the Agilent
81133A/81134A Pulse Generator, therefore this command tree is
redundant.
:STATus:PRESet
Syntax
:STAT:PRES
Form Event
Description This command
• clears all status group event- registers
• clears the error queue
• presets the status group enable, PTR, and NTR registers as
follows
:
Parameter
*RST value
Status Group Register Preset Value
OPERation ENABle 0000000000000000
PTR 0111111111111111
NTR 0000000000000000
QUEStionable ENABle 0000000000000000
PTR 0111111111111111
NTR 0000000000000000
QUEStionable:VOLTage ENABle 0111111111111111
PTR 0111111111111111
NTR 0000000000000000
QUEStionable:FREQuency ENABle 0111111111111111
PTR 0111111111111111
NTR 0000000000000000
QUEStionable:MONotony ENABle 0111111111111111
PTR 0111111111111111
NTR 0000000000000000
Vue de la page 69