Status Handling Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 71
:STATus:QUEStionable
This command tree accesses the QUEStionable status group.
The QUEStionable status group contains the summary bits from
the QUEStionable:VOLTage, :FREQuency and MONotony status
group.
The following commands are used to access the registers within
the status group.
:STATus:QUEStionable[:EVENt]?
Syntax
:STAT:QUES[:EVEN]?
Form Query
Description Reads the event register in the QUEStionable status group.
Parameter
*RST value
:STATus:QUEStionable:CONDition?
Syntax
:STAT:QUES:COND?
Form Query
Description Reads the condition register in the QUEStionable status group.
NOTE The Agilent 81133A/81134A Pulse Generator does not use this
condition register, therefore, this command is redundant.
Parameter
*RST value
:STATus:QUEStionable:ENABle
Syntax
:STAT:QUES:ENAB[?]
Form Set & Query
Description Sets or queries the enable register in the QUEStionable status
group.