SCPI Commands Reference SOURce Commands
64 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
[:SOURce]:VOLTage[0|1|2][:LEVel]
[:IMMediate]:TERM
Syntax
[:SOUR]:VOLT[0|1|2][:LEV][:IMM]:TERM[?]
Form Set & Query
Description Programs the termination voltage of the output signal for the
trigger output and the channels.
Parameter Numeric [uV|mV|V]
*RST value Trigger output (0): 0.0 V
Channels 1 and 2: 0.0 V
Specified limits Trigger output (0): –2.0 V … +3.0 V
Channels 1 and 2: –2.0 V … +3.0 V
Example Set Channel 1 termination voltage to 1 V.
:VOLT1:TERM 1V
[:SOURce]:VOLTage[1|2]:LIMit
[:AMPLitude]?
Syntax
[:SOUR]:VOLT[1|2]:LIM[:AMPL]?
Form Query
Description Reads the current setting of the amplitude limit. The result is only
valid if the “Limit to current levels” output mode is currently on
(
[:SOURce]:VOLTage[1|2]:LIMit:STATe ON
).
*RST value 100 mV
Example Read Channel 1 amplitude limit.
:VOLT1:LIM?