Agilent Technologies 81134A Manuel de service Page 68

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 104
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 67
SCPI Commands Reference Status Handling Commands
68 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
The following table shows the Agilent 81133A/81134A Pulse
Generator Status Handling Commands:
Command Parameter Description
:STATus
:OPERation
[:EVENt]?
Reads operation event register
:CONDition?
Reads operation condition register
:ENABle[?]
Numeric Sets/reads operation enable register
:NTRansition[?]
Numeric Sets/reads operation negative-
transition filter
:PTRansition[?]
Numeric Sets/reads operation positive-
transition filter
:PRESet
Clears and presets status groups
:QUEStionable
[:EVENt]?
Reads questionable event register
:CONDition?
Reads questionable condition
register
:ENABle[?]
Numeric Sets/reads questionable enable
register
:NTRansition[?]
Numeric Sets/reads questionable negative-
transition filter
:PTRansition[?]
Numeric Sets/reads questionable positive-
transition filter
:VOLTage
[:EVENt]?
Reads questionable voltage event
register
:CONDition?
Reads questionable voltage condition
register
:ENABle[?]
Numeric Sets/reads questionable voltage
enable register
:NTRansition[?]
Numeric Sets/reads questionable voltage
negative-transition register
:PTRansition[?]
Numeric Sets/reads questionable voltage
positive-transition register
:FREQuency
[:EVENt]?
Reads questionable frequency event
register
:CONDition?
Reads questionable frequency
condition register
:ENABle[?]
Numeric Sets/reads questionable frequency
enable register