Agilent Technologies DP235 Guide de l'utilisateur Page 118

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 117
118
Triggering Examples
Cross-Arming Trigger Examples
Cross-Arming Trigger Examples
The following examples use cross arming to coordinate measurements
between two separate analyzers within the logic analyzer or between
analyzers and the oscilloscope. The analyzers can be configured as
either a state analyzer and timing analyzer, or two state analyzers. It is
not possible to set both to timing.
You set up cross arming in the Arming Control menu (obtained by
selecting Arming Control in the Trigger menu). When coordinating
measurements between two instruments, you need to select Count
Time to correlate the measurements. When correlating measurements
between an analyzer and the oscilloscope you also need to set the
oscilloscope Trigger Mode to Immediate.
Vue de la page 117
1 2 ... 113 114 115 116 117 118 119 120 121 122 123 ... 627 628

Commentaires sur ces manuels

Pas de commentaire