Agilent Technologies DP235 Guide de l'utilisateur Page 426

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 425
426
Troubleshooting the Logic Analyzer
Analysis Probe Problems
Analysis Probe Problems
This section lists problems that you might encounter when using an
analysis probe. If the solutions suggested here do not correct the
problem, you may have a defective analysis probe. Refer to the User's
Guide for your analysis probe for test procedures. Contact your local
Agilent Technologies Sales Office if you need further assistance.
Target system will not boot up
If the target system will not boot up after connecting the analysis
probe, the microprocessor or the analysis probe may not be installed
properly, or they may not be making electrical contact.
Ensure that you are following the correct power-on sequence for the
analysis probe and target system.
1. Power up the analyzer and analysis probe.
2. Power up the target system.
If you power up the target system before you power up the analysis
probe, interface circuitry in the analysis probe may latch up,
preventing proper target system operation.
Verify that the microprocessor and the analysis probe are properly rotated
and aligned, so that the index pin on the microprocessor (such as pin 1 or
A1) matches the index pin on the analysis probe.
Verify that the microprocessor and the analysis probe are securely inserted
into their respective sockets.
Verify that the logic analyzer cables are in the proper sockets of the
analysis probe and are firmly inserted.
Reduce the number of extender sockets.
See Also
"Capacitive loading" in the previous section of this chapter.
Vue de la page 425
1 2 ... 421 422 423 424 425 426 427 428 429 430 431 ... 627 628

Commentaires sur ces manuels

Pas de commentaire