Logic Analyzer Reference
System Utilities
Update FLASH ROM field
The logic analyzer uses flash ROMs to store the operating system. The
analyzer you received should have an operating system in place and
should also include the operating system files on a flexible disk, but you
may occasionally need to update the operating system. Update FLASH
ROM updates the analyzer's operating system.
CAUTION: Updating flash ROM without the proper files will damage the logic analyzer
operating system.
1 Go to the System Utilities menu.
2 Select Update FLASH ROM.
The analyzer warns, "Selecting Continue Will Erase & Update Flash
ROMs." and waits for you to select Cancel or Continue. If you select
Continue, the analyzer will be reset. If you need to save the
measurement data, select Cancel.
3 Select Continue.
The screen goes black and the analyzer performs its power-up self
tests. It then displays a message listing the required files and provides
further instructions.
4 Insert the update disk into the flexible disk drive.
5 To search only the flexible disk drive, press the Done key. To
search the flexible disk drive and then the hard drive, press any
other key.