Using the Oscilloscope
Using the Oscilloscope
Using the Oscilloscope
This chapter covers the oscilloscope common menus and calibration.
This chapter covers:
calibrating the oscilloscope
oscilloscope common menus.
The oscilloscope circuitry requires an operational accuracy calibration
by the user or service department under any of the following
conditions:
at six month intervals or every 1000 hours
if ambient temperature changes more than 10°C from the temperature at
full calibration
to optimize measurement accuracy.
To test the oscilloscope circuitry against specifications (full
calibration), refer to chapter 3, Testing Performance in the “Agilent
Technologies 1670G-Series Logic Analyzers Service Guide.”